Sterilization Cycles

Pre-Programmed Sterilization Cycle Programs

The 5075 HSG autoclave includes 5 fixed sterilization programs, 2 test programs and 22 customizable sterilization programs.

 

Program
Load Type
Temp.
(oC)
Sterilization
Time (min)
Dry Time
 (min)
Total Time without drying
 (min) *
Unwrapped 134 Unwrapped instruments 134 4 1 24
Wrapped 134
Wrapped instruments
134
7
20 37
Unwrapped 121
Unwrapped delicate instruments
121
20
0 36
Wrapped 121
Wrapped delicate instruments
121
20
20 50
Prion 134
Prion
134
18
20 50
Leak Test
No load
-
-
- 31
Bowie & Dick Test PCD 134 3.5 2 21

 

* 5075 HSG cycle times are based on a 8kg load and  do not include drying. Cycle times based on warm chamber and may change according to load size.   Bowie & Dick Test is also suitable for Helix test device. Leak Test checks that the chamber under vacuum does not leak.

 