Description |
|
|---|---|
Standard Cycles |
Solid and glassware loads at 134°C or 121°C for delicate loads (plastics) Liquid loads and waste liquids at 121°C |
Standard Cycles added with optional features |
Cooling Coil + Compressed Air Fast cooling for liquid loads at 121°C |
Feature: Biohazard Air Filter |
During air removal all exhaust air is filtered through a 0.2 μm biological filter to prevent contamination of the laboratory Biohazard solid loads at 134°C Biohazard liquid loads at 121°C |
Feature: Vacuum Pump |
Pre-vacuum cycles Solid and glassware loads at 134°C or 121°C for delicate loads (plastics) Liquid loads and waste liquids at 121°C Air Leakage Test Cycle |
Feature: Vacuum Pump + Steam Generator |
Pre and Post vacuum cycles Hollow, porous and textile loads at 134°C Waste: hollow, porous and textile at 121°C Bowie & Dick steam penetration test at 134°C |
Media Processing Cycles |
Isothermal Processing: for preparing agar and other biological media with a temperature range from 60°C to 95°C that allows for gentle heating and cooling down of agar Holding Temperature: Special program with programmable holding temperature at the end of the cycle to prevent cooling of media |
Special Custom Cycles |
Tuttnauer is able to provide specifically customized cycles upon request. These may include material stress test, ageing test, varnish test, and others. Extended Sterilization Time: Special program with extended sterilization times up to 999 minutes Multiple Cycle (Material Stress Test): Special program to automatically run multiple cycles on the same load |
Print a hard copy of cycle data via optional printer
Store 200 cycles on your autoclave for reprinting or downloading to a PC
Easily transfer cycle data to a PC via USB
Connect to a local network for use with RPCR Software for remote monitoring of up to 8 autoclaves
Generate PDF reports with graphs and tables using optional RPCR Software
Multilevel Password protection
Overall Dimensions (WxHxD)
730mm x 1000mm x 540mm
Chamber Dimensions Ø x D
380mm x 690mm
Chamber Volume
85 Liter
Sterilization Temp. Range
105°C to 138°C
Voltage & Frequency
3 Phase
208V, 230V, 400V
50-60 Hz
Autoclave Weight
113 kg
2002/95/EC RoHS Directive
2006/95/EC Electrical equipment2004/108/EC Electromagnetic compatibility2006/42/EC Machinery Directive2002/96/EC WEEE DirectiveANSI / AAMI – ST55: 2010 Tabletop steam sterilizerEN 13060: 2004+A2: 2010 - Small Steam Sterilizer