Tuttnauer benchtop laboratory autoclaves are front loading autoclaves available in chamber sizes from 28 to 160 liters. EL models have an advanced multi-color control panel and a chamber made of 316L stainless steel. Tuttnauer laboratory autoclaves have been designed to provide high quality repeatable performance and accountability for a wide range of applications used in modern laboratories.
Each model has a number of optional added-value features which can be configured for fast cooling, efficient drying, biohazard and waste sterilization, Fo control, and more.
Benchtop lab autoclaves are designed to save space on your laboratory workbench.
Benchtop D-Line Models - Technical Data
D-Line Model
Chamber Dimensions ØxD (mm)
Chamber Volume (Liter)
External Dimensions WxHxD (mm)
2840 EL
280 x 400
28
530 x 440 x 630
3850 EL
380 x 500
65
720 x 540 x 765
3870 EL
380 x 690
85
720 x 540 x 940
5075 EL
500 x 750
160
860 x 740 x 1120
* External dimensions may change when an optional internal steam generator is added
Erlenmeyer Flasks (ml) Loading Capacity
Model
250
500
1000
2000
3000
5000
2840
1 x 9
1 x 4
1
0
0
0
3850
1 x 17
1 x 11
1 x 6
1 x 3
0
0
3870
1 x 23
1 x 17
1 x 10
1 x 5
0
0
5075
1 x 50 - 1 x 18
1 x 18 - 1 x 18
1 x 15
1 x 7
1 x 4
1 x 3
Schott-Duran Flasks (ml) Loading Capacity
Model
250
500
1000
2000
3000
5000
2840
1 x 12
1 x 9
1 x 5
0
-
0
3850
1 x 27
1 x 16
1 x 11
1 x 6
-
0
3870
1 x 36
1 x 23
1 x 18
1 x 9
-
0
5075
1 x 55 - 1 x 39
1 x 32 - 1 x 27
1 x 20
1 x 10
-
1 x 5
* Optional shelves are needed to accommodate some of the loading capacities mentioned above.
Baskets and Containers
Benchtop stainless steel wire baskets and containers in different sizes for all autoclave models.
Safety
Safety for personnel, autoclave and load are priority in the design, construction and operation of any Tuttnauer autoclave. Tuttnauer is committed to the highest industry safety standards and directives to ensure safety not only for your employees operating the autoclaves but also for your laboratory and the loads being sterilized.
A safety device prevents the operator from opening the door when the chamber is pressurized
Steam will not enter the chamber when the door is open
A cycle cannot start if the door is open or not properly locked
The door cannot unlock until liquid temperature reaches the predetermined end temperature
Two independent flexible PT100 temperature sensors to prevent over-boiling of liquids and explosions of bottles