TLC Laboratory Autoclaves Round chamber lab sterilizer

  1. Chamber volume of 160 liters
  2. Liquid sterilization with various cooling options
  3. Tips and Glass sterilization
  4. Instrument sterilization (wrapped or unwrapped)
  5. Biohazard and Waste sterilization
  6. Agar preparation

 

TLC-Line Advanced Laboratory Autoclaves

Tuttnauer laboratory autoclaves are designed to provide high quality repeatable performance and accountability for a wide range of applications used in modern laboratories:

Standard Autoclave Features

  • 7" Touch screen both sides
  • Automatic hinged doors
  • 18kW Built-in steam generator
  • Built-in vacuum pump
  • 316L Chamber
  • Coiled jacket
  • Fast cooling system
  • 304 Stainless steel housing
  • 2 Flexible temperature probes
  • Built-in printer
  • Built-in wheels
  • Base plate shelf
Each model has a number of additional optional features such as fast cooling, efficient drying, biohazard and waste sterilization, Fo control, specialized custom cycle programs, and more.

 

 


 


Technical Specifications

ModelsDoorsChamber VolumeChamber DimensionsExternal Dimensions
   Ø x D (mm)W x H x D (mm)

5075TLC-PVG-1A

5075TLC-PVG-2A

1 door

2 doors

160 Liters500 x 810

837 x 2027 x 1155

837 x 2027 x 1010

 

TLC Electrical Specifications (For all Models)

Voltage (V)380 (3- phase) / 230 (3-phase) / 230 (1- phase)*
Power18 kW
Frequency50-60 Hz
* requires additional 1-to-3 phase switch-box.
 

Schott Duran Flasks (ml) Loading Capacity
Standard base tray / Optional 2 Trays

Model250500100020005000
5075TLC Standard493623158
5075TLC
2 Trays
1056421104

Erlenmeyers Flasks (ml) Loading Capacity
Standard base tray / Optional 2 Trays

Model250500100020005000
5075TLC Standard36211483
5075TLC
2 Trays
67401183

Safety

Safety for personnel, autoclave and load are priority in the design, construction and operation of any Tuttnauer autoclave. Tuttnauer is committed to the highest industry safety standards and directives to ensure safety not only for your employees operating the autoclaves but also for your laboratory and the loads being sterilized.

  • A safety device prevents the operator from opening the door when the chamber is pressurized
  • Steam will not enter the chamber when the door is open
  • A cycle cannot start if the door is open or not properly locked
  • The door cannot unlock until liquid temperature reaches the predetermined end temperature
  • Two independent flexible PT100 temperature sensors to prevent over-boiling of liquids and explosions of bottles
Need more Information?
Brochure
Download the brochure and learn more about this autoclaves capabilities
Need help choosing?
Ready to place your order? We are here to assist you with any questions you may have.
Was this helpful?
Sorry about that
How can we improve it?
Thanks for your feedback!