Cycle Programs
5 pre-set programs, 2 test programs Bowie & Dick and Leak test, and 20 customizable programs.
Bowie & Dick test, also suitable for Helix test device, includes 3.5 min sterilization time without drying.
Leak test checks that the chamber under vacuum does not leak.
Cycle Program
|
Load Type |
Temp.
(oC) |
Sterilization
Time (min) |
Dry Time (min) |
Cycle Time (min) * |
Unwrapped 134 | Unwrapped instruments | 134 | 4 | 2 | 27 |
Wrapped 134
|
Wrapped instruments |
134
|
4
|
30 | 48 |
Unwrapped 121
|
Unwrapped delicate instruments |
121
|
20
|
2 | 38 |
Wrapped 121
|
Wrapped delicate instruments |
121
|
20
|
30 | 62 |
Prion 134
|
Prion |
134
|
18
|
30 | 63 |