The 50125GS autoclave is designed for Hospital CSSD, OR, and Medical Centers with limited budgets that need reliable sterile processing. Tuttnauer has maintained its high quality for hospital grade autoclaves, providing reliable, heavy-duty high-performance machines. The rounded chamber and coiled jacket reduce the manufacturing costs which allow for more affordable autoclaves.
Packed with all the features and reliability of a large capacity autoclave at a fraction of the size and cost.
More Affordable
Rounded chamber and coiled jacket reduce the manufacturing cost of the autoclave for facilities with limited budget
Simplified Design
Simplified design allows for low maintenance while maintaining hospital grade performance
Low Maintenance
Less parts and reduced complexity result in low long-term maintenance costs
Narrow Design
Designed for limited space. Narrow 84 cm width allow for more autoclaves in a limited area, reducing downtime risk
Convenient
1 Wall pass-through installation between non-sterilie and sterile rooms. Additional service area not required
Stainless Steel Piping
Piping is made of top-quality stainless steel to avoid corrosion and contamination of the sterilized loads
An optimized program for every load
The 50125GS comes with 8 pre-set and 20 customizable cycle programs, as well as 2 standard diagnostic tests.
Preset Cycles
Load Type
Temperature
Sterlization Time
Drying Time
Cycle Time*
Unwrapped instruments
134 °C
4 min.
1 min.
21 min.
Wrapped instruments
134 °C
7 min.
20 min.
32 min.
Unwrapped delicate instruments
121 °C
20 min.
1 min.
34 min.
Wrapped delicate instruments
121 °C
20 min.
20 min.
41 min.
Prion instruments
134 °C
18 min.
20 min.
47 min.
Leak Test
-
-
-
20 min.
Bowie & Dick Test
134 °C
3.5 min.
2 min.
25 min.
* Cycle times based on a 33kg load for 50125GS, excluding drying, based on warm chamber, and may change according to load. Bowie & Dick Test is suitable for Helix test device. Leak Test checks that the chamber under vacuum does not leak.
Test Cycle
Verify proper air removal from the autoclave chamber.
Leak Test Cycle
Verify the integrity of the pressure vessel and its plumbing on a daily or weekly basis.
Your data, where you need it
The 50125GS gives you easy access to its cycle data. Via USB, Ethernet or printer.
Your data, where you need it
The 50125GS gives you easy access to its cycle data. Via USB, Ethernet or printer.
7” Multi-color touch screen panel on both sides
Multiple access levels for added security
Graphical display of Temperature and Pressure trend graph
Stores the last 200 cycles in built-in memory
USB and Ethernet Ports
Multilingual interface (26 languages)
Technical specifications
Chamber Dimensions (Ø x L)
500mm x 1250mm
Chamber Volume
250 Liters
Overall Dimensions (WxHxD) 1 Door
837mm x 2027mm x 1515mm
Overall Dimensions (WxHxD) 2 Door
837mm x 2027mm x 1456mm
Basket Capacity
2 StU,
4 ISO
Tray Dimensions (WxHxD)
1 Door: 310mm x 21mm x 610mm, 2 Door: 471mm x 25mm x 610mm
EN 61010-1: 2010 Safety Requirements for measurement control and laboratory use
EN 61326-1: 2006 Electrical Equipment for EMC Requirements
EN 61010-2-40: 2005 Safety requirements for sterilizers used to treat medical materials
EN 17665-1: 2006 Sterilization of health care products - moist heat
Pressure Vessel
All pressure vessels and boilers are certified according to ASME code section I and section VIII, Div. I and PED 97/23/EC - Pressure Equipment Directive
Quality Standards
• ISO 9001:2008 Quality Management Systems
• EN ISO 13485:2012 Quality Management System - Medical Devices
Directives
• MDD 93/42/EEC amended in Directive 2007/47/EEC
• 2002/96/EC WEEE and 2002/95/EC RoHS
Need more Information?