Load Type |
Temperature |
Sterlization Time |
Drying Time |
Cycle Time* |
|---|---|---|---|---|
Unwrapped instruments |
134 °C |
4 min. |
1 min. |
21 min. |
Wrapped instruments |
134 °C |
7 min. |
20 min. |
32 min. |
Unwrapped delicate instruments |
121 °C |
20 min. |
1 min. |
34 min. |
Wrapped delicate instruments |
121 °C |
20 min. |
20 min. |
41 min. |
Prion instruments |
134 °C |
18 min. |
20 min. |
47 min. |
Leak Test |
- |
- |
- |
20 min. |
Bowie & Dick Test |
134 °C |
3.5 min. |
2 min. |
25 min. |
Test Cycle
Verify proper air removal from the autoclave chamber.
Leak Test Cycle
Verify the integrity of the pressure vessel and its plumbing on a daily or weekly basis.
7” Multi-color touch screen panel on both sides
Multiple access levels for added security
Graphical display of Temperature and Pressure trend graph
Stores the last 200 cycles in built-in memory
USB and Ethernet Ports
Multilingual interface (26 languages)
Chamber Dimensions (Ø x L)
500mm x 1250mm
Chamber Volume
250 Liters
Overall Dimensions (WxHxD) 1 Door
837mm x 2027mm x 1515mm
Overall Dimensions (WxHxD) 2 Door
837mm x 2027mm x 1456mm
Basket Capacity
2 StU,
4 ISO
Tray Dimensions (WxHxD)
1 Door: 310mm x 21mm x 610mm, 2 Door: 471mm x 25mm x 610mm
Voltage & Frequency
3 Phase (1ph option available)
230V, 400V
50-60 Hz
Steam Generator
18 kW