250 Liter chamber

50125GS Cost effective, hospital grade autoclave

The 50125GS autoclave is designed for Hospital CSSD, OR, and Medical Centers with limited budgets that need reliable sterile processing. Tuttnauer has maintained its high quality for hospital grade autoclaves, providing reliable, heavy-duty high performance machines. The rounded chamber and coiled jacket reduce the manufacturing costs which allow for more affordable autoclaves.

50125 GS Autoclave
Explore this product

Narrow, convenient and affordable

Packed with all the features and reliability of a large capacity autoclave at a fraction of the size and cost.
More Affordable
Rounded chamber and coiled jacket reduce the manufacturing cost of the autoclave for facilities with limited budget
Simplified Design
Simplified design allows for low maintenance while maintaining hospital grade performance
Low Maintenance
Less parts and reduced complexity result in low long-term maintenance costs
Narrow Design
Designed for limited space. Narrow 84cm width allow for more autoclaves in a limited area, reducing downtime risk
Convenient
1 Wall pass-through installation between non-sterilie and sterile rooms. Additional service area not required
Stainless Steel Piping
Piping is made of top-quality stainless steel to avoid corrosion and contamination of the sterilized loads

An optimized program for every load

The 50125GS comes with 5 pre-set and 22 customizable cycle programs, as well as 2 standard diagnostic tests
Preset Cycles
Load Type Temp. Sterlization Time Drying Time Cycle Time*
Unwrapped instruments 134 °C 4 min. 1 min. 21 min.
Wrapped instruments 134 °C 7 min. 20 min. 32 min.
Unwrapped delicate instruments 121 °C 20 min. 1 min. 34 min.
Wrapped delicate instruments 121 °C 20 min. 20 min. 41 min.
Prion instruments 134 °C 18 min. 20 min. 47 min.
Leak Test - - - 20 min.
Bowie & Dick Test 134 °C 3.5 min. 2 min. 25 min.
* La duración de los ciclos se basa en una carga estándar de 33 kg y no incluye secado. La duración de los ciclos se basa en que la cámara esté caliente y puede variar según el tamaño de la carga. La prueba Bowie & Dick también es adecuada para el dispositivo de prueba Helix.
 
Test Cycle
Verify proper air removal from the autoclave chamber.
 
Leak Test Cycle
Verify the integrity of the pressure vessel and its plumbing on a daily or weekly basis.

Your data, where you need it

The 50125GS gives you easy access to its cycle data. Via USB, Ethernet or printer.

Your data, where you need it

The 50125GS gives you easy access to its cycle data. Via USB, Ethernet or printer.
USB and Ethernet Ports
7” Multi-color touch screen panel on both sides
Multiple access levels for added security
Multilingual interface (26 languages)
Graphical display of Temperature and Pressure trend graph
Stores the last 200 cycles in built-in memory

Technical specifications

 

Technical specifications

Chamber Dimensions (Ø x L)
500mm x 1250mm
Chamber Volume
250 litros
Overall Dimensions (WxHxD)
1 Door
837mm x 2027mm x 1515mm
Overall Dimensions (WxHxD)
2 Doors
837mm x 2027mm x 1456mm
Basket Capacity
1 Stu
2 ISO
Tray Dimensions (WxHxD)
1 Door: 310mm x 21mm x 610mm
2 Doors: 471mm x 25mm x 610mm
Voltage & Frequency
3 phase (1ph option available)
208V, 230V, 400V 
50-60 Hz
Potencia
18 kW

Standards and directives

(a partial list)
Standards
EN285: 2006+A2:2009
ANSI/AAMI - ST 8: 2008
EN 61010-1: 2010 Safety Requirements for measurement control and laboratory use
EN 61326-1: 2006 Electrical Equipment for EMC Requirements
EN 61010-2-40: 2005 Safety requirements for sterilizers used to treat medical materials
EN 17665-1: 2006 Sterilization of health care products - moist heat
Pressure Vessel
All pressure vessels and boilers are certified according to ASME code section I and section VIII, Div. I and PED 97/23/EC - Pressure Equipment Directive
Quality Standards
  • ISO 9001:2008 Quality Management Systems
  • EN ISO 13485:2012 Quality Management System - Medical Devices
Directives
  • MDD 93/42/EEC amended in Directive 2007/47/EEC
  • 2002/96/EC WEEE and 2002/95/EC RoHS
Need more Information?
Brochure
Download the brochure and learn more about this autoclaves capabilities
Need help choosing?
Ready to place your order? We are here to assist you with any questions you may have.
Was this helpful?
Sorry about that
How can we improve it?
Thanks for your feedback!
PDF
Get Info